4.0 Operation

4.1 On / off switch

4.2 Suction system

!
Do not operate the suction system without a bacterial filter. Always use the splash guard, this will prolong the service life of the
bacterial filter. The guarantee will expire in the case of improper use!
Change the bacterial filter regularly. See details in chapter 5.0.
The hose attachment must be exchanged after each patient.

4.2.1 Control and display of the suction performance

Remove the handle from the support to activate the suction system. The suction system is automatically activated via the
installed light barrier. Set the desired vacuum via the controller. The end of the suction hose must be closed. Only in this case
can vacuum be built up.

4.2.2 Hose rinsing system

After each application rinse the suction hose thoroughly. Add the suction hose (without canula) with the suction nipple to the
suction attachment for the hose rinsing and soak up the rinsing fluid at least 30 sec.
14
After switching on the mains switch, all integrated and
connected devices are ready for operation.
at least 30
sec.
Display vacuum
Adjustment vacuum