Introduction ............3 Notes on operating instructions ......3 Intended use ............4 Function ..............4 Explanation of pictures and symbols ....5 Scope of supply ........... 5 For your safety ............6 Instructions for combination with other Medical Products .........6 Setting up and starting up .........7 Overview ..............
Scope and are therefore a must besides regular cleaning. ® Repair work and period tests may be carried out only by expert personnel authorised by ATMOS. By applying only original spare parts you will have the guarantee that operational safety, readiness for work and the value of your ATMOS Scope will be preserved.
1.2 Intended use Name ATMOS® Scope / ATMOS® Scope Basic Main function During an endoscopy procedure the use of the ATMOS® Scope / ATMOS® Scope Basic is indicated for the temporary application in the oral cavity up to the throat and in the nasal cavity. It enables the visualization of body orifices and body cavi- ties.
The product is MR unsafe - in areas with magnetic • The ATMOS Scope may be operated only in rooms ® resonance imaging, the product is unsafe. used for medical purposes, but not in areas subject to •...
3.0 Setting up and starting up Overview Fig. 1 Deflection lever Plug Pressure compensation valve with dust protection cap Plug cap Microphone (optional) Deflectable tip Button A Tip cover glass Button B Microphone interface Connection cable...
3.0 Setting up and starting up 3.2 Front view ATMOS Scope controller ® Fig. 2 Socket for connection cable ON/OFF switch 3.3 Rear view ATMOS Scope controller ® Fig. 3 BNC video outputs (composite/CVBS signal) 3.5 mm jack plug (Remote 1 signal) 3.5 mm jack plug (Remote 2 signal)
® • Install the driver from the enclosed CD ª After the installation is completed the ATMOS Scope will appear as a "USB 2820 Device" in the Device Manager. ® • Remove software CD from the drive. • Restart the PC.
Page 11
Strobo 21 LED operating instructions. ® Connecting the ATMOS Scope ® • Connect the ATMOS Scope handle with the controller, plugging the plug of connection cable into the socket for ® connection cable Application with the ATMOSoft / ATMOS Capture Suite ®...
3.0 Setting up and starting up 3.7 Tests Depending on the working distance (see technical data) the endoscopy image must be high-resolving, bright and clear. The following test sequence must be conducted on the device before processing and immediately before application. Testing the glass surfaces •...
Page 13
3.0 Setting up and starting up Testing the deflection mechanism • Slowly operate the deflection lever (Fig. 1), to check the function. • Check if full deflection is achieved (see chapter Technical Data). Limitations in deflection can indicate an endoscope defect. To avoid serious damage to the endoscope, only use the endoscope if the deflection works smoothly and without limitation. Risk for patients Risk for patients due to the use of products with sharp edges or damaged surfaces. •...
Never move the tip of the flexible endoscope against a resistance. The tip contains optical components, which may scratch or break when used incorrectly. 4.1.1 Switching on and adjusting the system Proper system operation cannot be guaranteed if the controller is switched on before the ATMOS Scope handle is connected. ® • Ensure, that the ATMOS Scope is connected to the controller before switching on the system.
4.0 Operation 4.1.2 Controller fine adjustment Generally the product does not need to be configured. However the user has the option to make fine adjustments. • Turn the controller off by pressing the ON/OFF switch • Press and hold either button A or B and turn controller ON. The Set-up menu will appear on the monitor. • Release button A or B •...
4.0 Operation 4.1.3 Functions of the buttons with ATMOSoft / ATMOS Capture Suite ® The ATMOSoft / ATMOS Capture Suite has separate operating instructions. ® Both cables must be plugged in Control box for connecting (see diagram). The Remote 1 in...
5.0 Cleaning and care 5.1 General instructions Adhere to national legal regulations, national and international standards and regulations as well as internal hygiene regulations for reprocessing. Reprocessing has to be carried out according to the manufacturer’s rules and instructions. Only use recommended agents and procedures.
• Only reconnect the system to the mains supply when all cleaned parts are completely dry. 5.2.2 Cleaning and disinfection: ATMOS Scope handle with flexible endoscope part ® Manual cleaning with Tristel Wipes System Please follow the manufacturer's information and operating instructions.
Page 19
5.0 Cleaning and care Stage I • Fully immerse the product in the cleaning solution. Ensure that all accessible surfaces are moistened. • Clean the product while it is in the cleaning solution, using a soft cloth or a suitable cleaning brush, until all visible residues have been removed from the surfaces.
• Position the endoscope according to the instructions of the manufacturer in the AFER. • Connect tube of the AFER for the leakage test to the pressure compensation valve of the ATMOS Scope handle. If ® necessary, use adapters. The validation that the product can be cleaned and disinfected has been conducted using the automatic endoscope reprocessors of Wassenburg Typ WD 440 in standard programme.
5.0 Cleaning and care 5.4 Sterilization 5.4.1 General instructions Do not autoclave the product. Before sterilization ensure that the endoscope has no restrictions in use. See chapter Start of Operation (section Tests). Ensure that the sterilization agent can reach all surfaces. The implementation of the sterilization lies within the responsibility of the user in order to attain the required degree of sterilization.
Page 22
• Clean and disinfect the ATMOS Scope manually or mechanically. ® • Allow the ATMOS Scope to dry thoroughly before you sterilise it in the STERRAD sterilizer. Humidity can lead to a ® termination of the sterilisation cycle. • Tighten the pressure compensation cap.
ATMOS recommends: Work should be carried out by an authorised ATMOS service partner. This ensures that the repairs and testing are carried out professionally, that original spare parts are used and that warranty claims remain unaffected.
Camera USB driver is not installed on the PC Install camera USB driver via software CD monitor ATMOS Scope is not recognized by USB port. USB port via Windows/control panel/power settings ® Perhaps the USB port is turned off must be permanently activated...
9.0 Technical data Voltage 115-230 V~ ± 10 %; 50/60 Hz Current consumption Max. 0.5 A Power consumption 30 VA Fuses 2 x T 630 mA L / 250 V Probe diameter 3.8 mm Probe length 300 mm Deflection 2 x 135° Direction of view 0°...
10.0 Disposal • The ATMOS Scope does not contain any hazardous materials. ® • The housing material is fully recyclable. • Device and accessories must be decontaminated prior to disposal. • Pay attention to a careful separation of the different materials.
• The use of other accessories, other transducers and cables than stated may lead to an increased emission or a reduced interference immunity of the equipment or system. 11.1 Guidelines and Manufacturer's Declaration - Emissions The ATMOS® Scope is intended for use in the electromagnetic environment specified below. The customer or user of the AT- MOS® Scope should ensure that it is used in such an environment. Emissions Test Compliance Electromagnetic Environment - Guidance RF Emissions acc.to CISPR 11...
Page 30
AC mains voltage prior to application of the test level. 11.3 Guidelines and Manufacturer's Declaration - Immunity The ATMOS Scope is intended for use in the electromagnetic environment specified below. The customer or user of the ® ATMOS Scope should ensure that it is used in such an environment. ® Immunity Test IEC 60601- Test Compliance Level Electromagnetic Environment - Guidance...
Page 31
Scope is used exceeds the above compliance level, the ® ATMOS Scope is to be observed to verify the intended use. If abnormal performance characteristics are noted, additional ® measures might be necessary, e. g. a changed arrangement or another location for the ATMOS Scope.