10.0 Technical data at complete equipment
Please note the technical specifications of the single devices in the separate operating instructions!
Voltage
Current consumption
Power consumption
Fuses
Suction system standard
• Power density (free flow)
• Vacuum
• Secretion canister
Suction system professional
• Power density (free flow)
• Vacuum
• Secretion canister
Compressed-air system
• Power density (free flow)
• Pressure
• Sprayer
Ear irrigation basic
• Water temperature
Ear irrigation professional (Hygrotherm)
• Water temperature
• Flow
Thermal nystagmus stimulation (Variotherm)
• Water temperature
• Flow
• Timer setting
• Interface
Instrument heating
• Temperature
Light module economy
• LED power supply
LED cold light module
• Number of channels
• Light intensity
• Colour temperature
Operating time
Protective earth conductor resistance
Earth leakage current
Enclosure leakage current
Patient leakage current
Ambient conditions for transport/storage
• Temperature
• Humidity without condensation
• Pressure
230 V~ ± 10 %; 50/60 Hz
Max. 8.5 A (230 V~)
Max. 2,300 VA
2 x T 10 A / 250 V (f. 230 V~) 1 x M 15 A / 250 V (f. 100 V~, 115 V~, -127 V~)
40 l/min
-91 kPa (-910 mbar or 682,5 mmHg) @NN, optionally infinitely variable
DDS 1,0 l
55 l/min
-95 kPa (-950 mbar or 712,0 mmHg) @NN, optionally infinitely variable
DDS 1,0 l
20 l/min
Min. 220 kPa, infinitely variable
For liquid, oily and powdery materials
37.5 °C +/- 2 °C
Adjustable 36 - 38 °C +/-0.5 °C
Max. 450 l/min (adjustable on the handle)
3 adjustable levels from 20 – 47 °C +/-0.5 °C
Max. 450 l/min (adjustable on the handle)
1 – 99 s
Nystagmograph triggering
Approx. 37 °C
700 mA regulated
2
Min. 200 klx
5500 K +/- 10 %
Continuous operation
Max. 0.1 Ω
Max. 0.5 mA
Max. 0.1 mA
Max. 0.1 mA
-10...+50 °C
30...95 %
500...1060 hPa
Special voltage:
100 V~ ± 10 %; 50/60 Hz
115 V~ ± 10 %; 50/60 Hz
127 V~ ± 10 %; 50/60 Hz
Max. 19.0 A (100 V~)
Max. 16.5 A (115 V~)
Max. 15.0 A (127 V~)
39