4.0 Operation

4.6.2 Adapt / deposit sprayer

Adapt:
1
2
• Handle the sprayer carefully to avoid injuries.
• Before using the sprayer bottle, check the durability of the included medications.
• Clean all parts of the sprayer regularly! Information can be found in chapter 5.0 "Cleaning" and chapter 6.0 "Hygiene plan"!

4.6.3 Ear irrigation module compressed air (REF 530.2080.0)

 Adjusting lever
 Vent valve

4.7 LED-Light system

Insert () the handle onto the
medication bottle.
Engage, then lift medication
bottle with the handle from
the support ().
By pulling the trigger,
compressed-air will be
applied into the sprayer ().
3
2
1
On / off LED
LED port (for light source or headlight)
LED port (for light source or headlight), optional connection for nystagmus
binoculars
Deposit:
1
To start the ear irrigation, close the vent valve () and
press the adjusting lever (). To terminate ear irrigation,
release the adjusting lever () and the vent valve (). Ear
irrigation stops immediately!
Prior to every use the glass bottles must be checked for
!
intactness (cracks, chippings on the winding, etc.). Broken
glass vials may not be used for ear irrigation.
Insert the medication bottle
top down in the support.
2
Press activator () and pull
handle () backwards.
23