6.0 Hygiene Plan
What
Parts to be reprocessed
Instrument management
ENT instruments
Instrument bowl
Instrument bowl with cover
Visualization
ATMOS
®
ATMOS
®
ATMOS
®
Flexible Endoscope
Rigid endoscope
Laryngoscope
Light cable
Light grip
Microscope
Headlight
Radiofrequency surgery
ATMOS
®
Ergonomic handles
Bipolar tweezers
Bipolar electrode
Bipolar electrode cable
Neutral electrode
Neutral electrode cable
ENT electrodes
Surfaces
Housing
Roller shutter
System frame
Drawers
Writing surface
Instrument deposit
Mirror preheater
Tongue patches and swab
dispenser
Waste disposal
Instrument tray
Recommended disinfectants
Surface disinfection for coated surfaces:
3)
• Green & Clean SK (ATMOS)
• Dismozon
®
plus (Bode Chemie)
• Kohrsolin
FF (Bode Chemie)
®
• Perform
(Schülke & Mayr)
®
• Terralin
®
Protect (Schülke & Mayr)
Other surfaces:
• Dismozon
plus (Bode Chemie)
®
• Kohrsolin
FF (Bode Chemie)
®
• Mikrobac
®
forte (Bode Chemie)
• Perform
(Schülke & Mayr)
®
• Terralin
Protect (Schülke & Mayr)
®
• Surface disinfectant FD 312 (Dürr Dental)
Important information
Wipe cleaning and disinfection:
All surfaces have to be wiped with a clean
(disposable) wipe which is damped with
disinfectant solution. The entire surface has
to be wiped thoroughly and may not be dried
afterwards.
32
How
C
D
Cleaning
Disinfection
X
X
2,4,5
X
X
4
X
X
4
Cam 21 / 31
X
X
3
Strobo 21 LED
X
X
3
LS 31 LED
X
X
3
X
X
1,7,8
X
X
1,7,8
X
X
1,7,8
X
X
3
X
X
3
X
X
3
X
X
3
RS 221 (surface)
X
X
3
X
X
1,2,4.5
X
X
1,2,4.5
X
X
1,2,4.5
X
X
1,2,4.5
X
X
1,2,4.5
X
X
1,2,4.5
X
X
1,2,4.5
X
X
3
X
X
3
X
X
3
X
X
3
X
X
3
X
X
3
X
X
3
X
X
3
X
X
3
X
X
3
4)
Manual disinfection of instruments:
• Korsolex
med AF (Bode Chemie)
®
• Korsolex
basic (Bode Chemie)
®
• Korsolex
plus (Bode Chemie)
®
• Korsolex
®
extra (Bode Chemie)
• neodisher
Septo MED (Dr. Weigert)
®
• neodisher
Septo 3000 (Dr. Weigert)
®
• Sekusept
®
PLUS (Ecolab)
• Sekusept
aktiv (Ecolab)
®
• Gigasept
Instru AF (Schülke & Mayr)
®
• Gigazyme
(Schülke & Mayr)
®
• Gigasept FF neu (Schülke & Mayr)
Automatic disinfection of instruments:
5)
• Dismoclean
24 Vario (Bode Chemie)
®
• Dismoclean
28 alka med (Bode Chemie)
®
• Dismoclean
®
twin basic/twin zyme (Bode Chemie)
• neodisher
FA (Dr. Weigert)
®
• neodisher
MediClean forte (Dr. Weigert)
®
• Thermosept
alka clean forte (Schülke & Mayr)
®
• Thermosept
®
RKN-zym (Schülke & Mayr)
Please observe the manufacturer's operating
1)
instructions.
Preferred: machine cleaning and disinfection in the
2)
washer disinfector
6)
Material dimensionally stable at 134°C
Notices
S
Sterilization
Immerse instruments into solution immediately after use,
complete wetting is required, air must be removed from
any cavities, after the contact time instruments must
X
be rinsed with water, have to be dried and sterilised
afterwards. Please also observe the ATMOS operating
instructions for ENT instruments.
Cleaning and disinfection (manual)
Cleaning with a brush; disinfection (manual)
Wipe cleaning and disinfection
Wipe cleaning and disinfection
Wipe cleaning and disinfection
Immediate pre-cleaning after application
X
1
Immediate pre-cleaning after application
X
1
Immediate pre-cleaning after application
X
1
Wipe cleaning and disinfection
Wipe cleaning and disinfection
Wipe cleaning and disinfection
Wipe cleaning and disinfection
Wipe cleaning and disinfection
Wipe cleaning and disinfection
X
1
Immediate pre-cleaning after application;
X
1
Cleaning and disinfection (manual or automatic);
X
1
Use of enzymatic detergents
Immediate pre-cleaning after application;
X
1
Cleaning and disinfection (manual or automatic);
X
1
Use of enzymatic detergents
Immediate pre-cleaning after application;
X
1
Cleaning and disinfection (manual or automatic);
X
1
Use of enzymatic detergents
Wipe cleaning and disinfection
Wipe cleaning and disinfection
Wipe cleaning and disinfection
Wipe cleaning and disinfection
Wipe cleaning and disinfection
Wipe cleaning and disinfection
Wipe cleaning and disinfection
Wipe cleaning and disinfection;
Daily or when refi lling
Wipe cleaning and disinfection;
Daily or when refi lling
Wipe cleaning and disinfection;
Daily or when refi lling
7)
Endoscopes - manual disinfection:
 Helipur
H plus N (BBraun)
®
 Helix
Ultra (BBraun)
®
 Korsolex
Basic (Bode Chemie)
®
 neodisher
®
MediClean forte (Dr. Weigert)
 Sekusept
aktiv (Ecolab)
®
Endoscopes - automatic disinfection:
8)
 Korsolex
Basic (Bode Chemie)
®
 neodisher
®
MediClean forte (Dr. Weigert)
 Gigasept
FF neu (Schülke & Mayr)
®
 Endozime
AW Plus (Ruhof)
®
 ADAPTACLEAN™ (ASP)
For concentrations, contact time, temperature,
material compatibility, please see the relevant
information from the manufacturer.
Wrong concentration of disinfectants
may lead to material damage!
When
Who
Qualifi ed and trained
staff who are familiar
with reprocessing.
(Please fi ll in the re-
sponsible person, use
a water-based over-
head marker)
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
The above stated hygiene requirements are based on the regulations
according to the Medical Devices Act, the Medical Devices Operator
Ordinance, §18 IfSG and the recommendations of the Robert Koch Institute.
Defi nition of the required reprocessing steps result from the
recommendations of the Robert Koch Institute: „Requirements for the
reprocessing of medical products", from Robert Koch Institute. The medical
products were categorised in the risk groups uncritical, semi-critical
and critical. The reprocessing measures mentioned in this cleaning and
disinfection plan are a recommendation of ATMOS MedizinTechnik. Any
additional reprocessing measures are at the operator's discretion.
All the recommended disinfectants which are stated herein are listed
disinfectants (VAH/RKI) and have been tested on their suitability of use on
the ATMOS
S 61 Servant. ATMOS MedizinTechnik cannot be hold liable for
®
any damage caused by wrong concentration of the disinfectants or by the
application of any other disinfectants.
Patients with suspicion of a clinical disease or who developed a
transmissible spongiform encephalopathy (CJK, vCJK, etc.) have to be
treated at facilities which are able to provide for the necessary preventive
measures against infection. The reprocessing of the reusable instruments
and material may only be performed at facilities which have an externally
certifi ed QM Management acc. to DIN EN ISO 13485.
The Medical Devices Act, IfSG, the RKI directives, BGR 250 and TRBA 250
always have to be considered.
ATMOS MedizinTechnik GmbH & Co. KG
Ludwig-Kegel-Str. 16  79853 Lenzkirch/Germany
Phone +49 7653 689-0  Fax +49 7653 689-190
info@atmosmed.de  www.atmosmed.de
GA3GB.110100.0
2017-12 Index: 14