4.0 Operation

4.4.3 Activation of the Variotherm

Remove the handle from the support. The Variotherm is
automatically activated via the installed light barrier. The
adjustment control of the Variotherm requires ca. 20 s, until
the desired temperature is reached. Check the temperature
indication! As long as the temperature of the actual value
deviates from the target value, the LED button „set" flashes.
When the target and actual value correspond, the LED lights
up permanently. Now you may begin with the irrigation of the
auditory canal.
The last active temperature level is selected.
4.4.4 Performing the thermal nystagmus stimulation
Warning: May only be used with the ear irrigation bowl.
The hose attachment must be exchanged after each patient.
!
Remove the handle  select temperature and adjust if necessary  if required set time  press  (flashing approx. 10 sec.)
 when the Variotherm is ready for use a beep can be heard  hold nozzle in working position and activate trigger  perform
nystagmus stimulation  after the time expires a beep sounds again

4.4.5 Time setting (1-99 sec.)

Remove the handle.
8
Press and hold
2
set the desired stimulation time.
When selecting the respective temperature, the setting from the
previous time will be adopted automatically.
Now and again check the time setting of the timer.

4.4.6 Temperature setting

You can apply the Variotherm at temperatures from 20°C to 47°C.
The minimum temperature is defined by the temperature of the
tap water.
Setting at 37°C:
Press
5
for the temperature.
Warm stimulus (38 - 47°C):
Remove the handle.
Keep
pressed. Set the desired temperature with
6
Release.
When selecting the button
will be adopted automatically.
Cold stimulus (20 - 36°C):
Remove the handle
Keep
7
pressed. Set the desired temperature with
Release.
When selecting the button
will be adopted automatically.
Heating off:
Press the active temperature level on the button "Temperature" to
switch the heating on and off.
20
.
2
6
the previous temperature setting
2
.
the previous temperature setting
7
6
7
Ear irrigation bowl
Handle thermal
nystagmus stimulation
On / off switch
nystagmus binoculars
Connection
nystagmus binoculars
2
8
2
8
7
6
5