Thermal Nystagmus Stimulation, Variotherm; Overview; Switching On - Atmos S 61 Operating Instructions Manual

Servant ent workstation, ent treatment unit
Table of Contents

Advertisement

4.0 Operation

4.4 Thermal nystagmus stimulation, Variotherm

4.4.1 Overview

Front view
9
Display of stimulation time (double-digit, resolution 1s)
1
Adjustment button up or down (for temperature or
2
seconds setting)
Temperature indicator (double-digit resolution 1°C),
3
actual value indication
Maintenance indicator
4
Irrigation handle
3
7
5
11

4.4.2 Switching on

Attention: Make sure the water supply is guaranteed!
• When the device is switched on, a segment-test is performed. For a short period of time all LEDs will be activated. Visually
check the function of the LEDs during this time.
• Prior to the caloric measurements check the water pressure at the manometer in the service compartment! A target pressure
of 2 bar - max. 2.2 bar must be available. Danger by overpressure!
1
2
3
8
7
6
8
6
10
9
1
4
2
5
Button for selecting the irrigation level (37°C)
5
Button for selecting the warm stimulation level (e.g. 44°C)
6
Button for selecting the cold stimulation level (e.g. 30°C)
7
Button for adjusting the stimulation period
8
Button to Start/Stop the stimulation (150 ml/min)
9
Handle
1
Connection for double-barrelled hose
2
Three-hole gasket
3
Adjusting screw
4
Release lever
5
Valve plunger
6
Jet connection
7
Splash guard
8
Sealing cap
9
Double-barrelled hose
10
Set screw
11
4
19