12.0 Notes on EMC

• Medical electrical equipment is subject to special precautions with regard to EMC and must be installed acc. to following
EMC notes.
• Portable and mobile HF communication facilities can influence medical electrical equipment.
• The use of other accessories, other converters and cables than stated may lead to an increased emission or a reduced
interference immunity of the equipment or system.
12.1 Guidelines and Manufacturer´s Declaration - Emissions
The ATMOS
S 61 Servant workstation is intended for use in the electromagnetic environment specified below. The customer or
®
user of the ATMOS
S 61 Servant workstation should ensure that it is used in such an environment.
®
Emissions Test
RF Emissions acc.to CISPR 11
RF Emissions acc. to CISPR 11
Harmonic emissions according to
IEC 61000-3-2
Voltage fluctuations/flicker according
to IEC 61000-3-3
12.2 Guidelines and Manufacturer's Declaration - Immunity
The ATMOS
S 61 Servant ENT workstation is intended for use in the electromagnetic environment specified below. The
®
customer or user of the ATMOS
Immunity Test
Electrostatic discharge
(ESD) according to IEC
61000-4-2
Fast electrical transient/
burst IEC 61000-4-4
Surges IEC 61000-4-5
Magnetic field at power
frequency 50/60 Hz acc.
to IEC 61000-4-8
Compliance
Group 1
Class B
Class A
Corresponds
S 61 Servant ENT workstation should ensure that it is used in such an environment.
®
IEC 60601- Test Level
± 6 kV Contact
± 8 kV Air
± 2 kV Mains
± 1 kV I/Os
1 kV
Common
2 kV
Differential
3 A/m
Electromagnetic Environment - Guidance
The ATMOS
S 61 Servant workstation uses RF energy only
®
for its internal function. Therefore, its RF emissions are very
low and are not likely to cause any interference in nearby
electronic equipment.
The ATMOS
S 61 Servant workstation is suitable for use
®
in all establishments, including domestic and those directly
connected to the public low-voltage power supply network that
supplies buildings used for domestic purposes.
Compliance Level
± 6 kV Contact
± 8 kV Air
± 2 kV Mains
1 kV
Common
2 kV
Differential
Inapplicable
Electromagnetic Environment - Guidance
Floors should be wood, concrete, or ceramic
tile. If floors are synthetic, the relative
humidity should be at least 30 %.
Mains power quality should be that of a
typical commercial or hospital environment.
Mains power quality should be that of a
typical commercial or hospital environment.
Power frequency magnetic fields should
be that of a typical commercial or hospital
environment.