8.0 Troubleshooting

Description
Electric power supply
No function, main switch is active,
control light does not light up
Suction
Low or no suction rate, but vacuum
gauge indicates vacuum
Low or no suction rate, but vacuum
gauge indicates few or no vacuum
Suction pump does not switch on or
off
No suction, but force pump is
operating
Compressed air
Compressed air pump does not
switch on or off
No compressed air, but suction
pump is operating
No adequate pressure build-up
Sprayer does not work
Mirror quick heater
Heater does not switch on
Insufficient or no heater power
Automatic secretion canister evacuation
Canister will not be drained when
filled above the sensor limit
Pump does not activate when
suction hose is returned
Effluent pump does not activate after
switching off the draining pump
Effluent pump does not switch off
Effluent pump does not start
Cause
No voltage to power plug
Cable/plug defect
Suction hose is clogged
Active oversuction protection
Filter moistened/blocked
System leaky
Suction hose snapped off
Vacuum controller is open
Secretion penetrated the suction
pump
Draggled/defect light barrier in
suction hose holder
Positions of handles have been
mixed up
Draggled/defect light barrier in
compressed air holder
Positions of handles have been
mixed up
Leaky connections
Sprayer is clogged
Switch or control unit defect
Single or all three heater coils
defect, no glow
Sensor does not detect filling level
Pump defect
Pump defect
Effluent pipe blocked
Electrodes in the effluent canister
are dragged and thereby hot-wired
Device is fed with demineralised
water via water connection
Measure
Check house fuse, possibly it burnt through
Exchange cable/plug
Clean suction hose
Check oversuction protection
Replace filter
Check all joints, new assembly of the system
Remove breaks, pass hose a different way
Set regulator to higher/maximum vacuum
Inform the service department
Clean both light barriers openings in the
suction hose holder, if necessary inform service
department
Insert handles in the right holder
Clean both light barriers openings in the
compressed air holder, if necessary inform
service department
Insert handles in the right holder
Check joints, if necessary inform service
department
Clean sprayer, clean ventilation borehole in
sprayer head
Inform service department to exchange the
switch or control unit
Change heater coil
Clean canister thoroughly
Inform the service department
Inform the service department
Remove blockage
Clean effluent canister or let electrodes be
cleaned
Please use drinking water