4.0 Operation
4.4.7 Adjusting the flow quantity
Prior to stimulation the set screw
upwards position.
2 Adjustment possibilities:
• High: Flow quantity for ear irrigation
• Low: Flow quantity for stimulation of the vestibule
Prior to the ear irrigation the set screw is screwed downwards,
inserted into the handle and then locked into position by
continuing to screw it downwards.
For fine adjustment of the flow the set screw
in or out.
) Periodically check the water flow for the stimulation of the
vestibular organ and readjust, if necessary.
4.4.8 Information about the service display of the ATMOS
The ATMOS
S 61 Servant workstation has a service display on the front panel.
®
The display consists of a red and a yellow LED.
After switching on the module the electronics goes through a test cycle. The following indications are possible:
Immediately after start-up the LED test starts.
Start-up
Subsequently a functional and aggregate test follows. For this purpose six flashing signals light up (time-ratio light up/pause,
500 ms/500 ms).
Operating time pressure pump
Operating time suction pump
Operating time suction pump 2
Operating time hose pump
Operating time drainage pump
Annual maintenance
Battery voltage
Possible maintenance cases:
Operating time pressure pump
Operating time suction pump
Operating time suction pump 2
Operating time hose pump
Operating time drainage pump
Annual maintenance
Battery voltage
If one of the above maintenance cases occurs, please contact your local ATMOS
11
must be locked in the
must be screwed
4
red LED
yellow LED
Lights up
Lights up
First light up
Second light up
Third light up
Fourth light up
Fifth light up
Sixth light up
Seventh light up
First light up
Second light up
Third light up
Fourth light up
Fifth light up
Sixth light up
Seventh light up
11
4
S 61 Servant up from software version V 1.2
®
red LED
yellow LED
Beep
Remarks
Both LEDs light up simultaneously
Yes
during switch on (duration: 500
ms).
No
o.k.
No
o.k.
No
o.k.
No
o.k.
No
o.k.
No
o.k.
No
o.k.
Yes
Critical operating time
Yes
Critical operating time
Yes
Critical operating time
Yes
Critical operating time
Yes
Critical operating time
Yes
Annual maintenance is due
Yes
Low battery voltage
Service partner.
®