Atmos S 201 Thorax Operating Instructions Manual

Atmos S 201 Thorax Operating Instructions Manual

Hide thumbs Also See for S 201 Thorax:

Advertisement

English
312.1000.B
2014-08 Index: 33
ATMOS
S 201 Thorax
®

Advertisement

Table of Contents
loading

Summary of Contents for Atmos S 201 Thorax

  • Page 1 English ATMOS ® S 201 Thorax 312.1000.B 2014-08 Index: 33...
  • Page 2: Table Of Contents

    Notes on operating instructions ........3 General information on cleaning and disinfection ..21 Intended use ..............3 Cleaning the device surface ......... 21 Contraindication ATMOS S 201 Thorax ......3 ® Warnings ................ 3 Precautions ..............4 Maintenance and service ........... 22 Function .................
  • Page 3: Introduction

    Specifi cation of the main function: The digital drainage suction systems ATMOS S & E 201 Thorax are devices for mobile thoracic drainage. They are intended for short-time (<30 days) application on human beings. They are portable, mains-independent and have an electronic monitoring sy- stem with optical and acoustic status display.
  • Page 4: Contraindication Atmos ® S 201 Thorax

    The patient should be regularly monitored according to internal hospital guidelines. ● ATMOS S 201 Thorax is a medical device that requires special safety precautions and must be installed and ® placed in operation in accordance with the attached EMC information. Portable and mobile RF communication devices (mobile telephones) can affect the drainage system.
  • Page 5: Precautions

    ● During set up, an incorrectly positioned drainage system The ATMOS S 201 Thorax is a particularly useful small suction and patient tube may impair fluid and air evacuation. unit. The device is operated by an electromotive, maintenance- Total obstruction of fluid and air evacuation can cause free piston pump.
  • Page 6: Explanation Of Pictures And Symbols

    Pictures contained in this manual Warning,  Important information special diligent notice ! Operating keys / symbols of the ATMOS S 201 Thorax Application part type CF, defibrillator-proof, recovery time: 10 sec ON = short keypress (supply, connection to Serial number...
  • Page 7: Scope Of Supply

    Scope of supply ● Prior to dispatch, this ATMOS device was subjected to an extensive functional test and has been carefully packed. Nevertheless, please compare the contents of the shipment on completeness immediately upon receipt (see delivery note). Hose set...
  • Page 8: For Your Safety

     The environmental conditions given in the Technical Specifications (section 9.0) must be observed.  The ATMOS S 201 Thorax may only be used by trained specialist staff. The control panel should always be clearly in view for the ...
  • Page 9: Setting Up And Starting Up

    Mains voltage: 100-240 V; 50/60 Hz ● The device can only be operated with the special Fuse: 1 x T 1A, L ATMOS container, therefore please take care to have at least one spare container available. ● USA/ Canada: Use only „Hospital Grade“ mains plugs and cords with remark: Grounding reliability can only be achieved when the equipment is connected ...).
  • Page 10: Functional Test

    Setting up and starting up Functional test   Mounting of the ATMOS S 201 Thorax as described  under Fig. 2.   Before use the system must always be tested regarding any leakage.  Slip on the test hose set to the device and connect the ...
  • Page 11: Operation

    Fig 7 Hose Rinsing An automatic hose rinsing function will operate periodically. ● Only use the original ATMOS This rising function is designed to draw any secretion or single-use secretion container! fluid build-up in the measurement hose into the suction hose and subsequently the secretion container.
  • Page 12: Container

    Then move the container towards the housing until it clicks into place (Fig 10). The ATMOS S 201 Thorax is again ready for use. Reopen the clamp at the suction hose.
  • Page 13: Connect Hose Set

      Fig 13 Put the ATMOS S 201 Thorax onto the support. Slightly pull out the upper button  and rotate it by 90° until the two connectors  are at the level of the holes  and then...
  • Page 14: Information On The Display

    0,25 l/min (blue) from 10 to 12: 3,30 l/min (red) Fig 15 Display functions 4.9.1. Switch on Switch on the ATMOS S 201 Thorax with the I/O key. Fig 16 The start screen is shown for 2 seconds. 4.9.2 Adjusting the vacuum Adjust the target vacuum with the keys▼▲.
  • Page 15: Keylock

    Then move the container towards the housing until it clicks into place. The ATMOS S 201 Thorax is again ready for use. Start up the pump by pressing the I/O key. Reopen the clamp at the suction hose.
  • Page 16: Switch Off

    ▲▼ keys. If all settings are made, exit the settings menu in depressing the menu item „exit“ and confirm with the enter key  on the display. Your settings are also saved after switch off of the ATMOS S 201 Thorax. Fig 22 Enter/ Return...
  • Page 17: Alarm

    Before the device can be used, the correct function of the optical and acoustic display must be checked by the user. Perform the following steps: Attach a canister on the ATMOS S 201 Thorax. Connect a hose to the unit.
  • Page 18: Alarm History

    Operation 4.11.1 Alarm history You can enter the settings menu by a long keypress of the  button (3 sec). The first dialogue field is the field „language“. By pressing the ▲ or ▼ key you can enter the dialogue field „alarm history“ on page 2.
  • Page 19: Insertion And Removal Of The Micro Sd Card

    Operation 4.13 Carrying strap Fix the carrying strap with the velcro fastening to the handle, adjust the requested length of the carrying strap and put the device on. Fig 29 4.14 Insertion and removal of the micro SD card Insert the micro SD card upright and with the face up. Fig 30 Insert the micro SD card into the opening until it locks in place with a „click“.
  • Page 20: Monitoring The Progression Of Therapy Via Micro Sd Card

    Prior to use the device, the nurse call system must be checked (service menu item). The ATMOS S 201 Thorax can be connected to the nurse call system via the nurse call interface (NCI). A stereo phone jack 3.5 mm is required. In the case of an alarm the alarm signalling is transferred to the nurse call system.
  • Page 21: Cleaning

    ● The unit itself can be wiped off with a moist (not wet) cloth. ● The surfaces of the ATMOS S 201 Thorax can be cleaned / disinfected with products of the following active irgredients: QAV (quartanary ammonium compounds) ...
  • Page 22: Maintenance And Service

    ● Maintenance or opening and repair of the ATMOS authorised by ATMOS. The device should at least be E 201 Thorax may only be carried out by ATMOS or a reprocessed every 12 months. specialist authorised by ATMOS. In this case, attention...
  • Page 23: Trouble Shooting

    Device must be checked by the authorised service Service is required. Contact ATMOS Service hardware / software fault Service is required. If necessary Contact ATMOS Service. During a restart the restart without mains connection. error message appears every 30 seconds.
  • Page 24: Accessories, Consumables And Spare Parts

    Accessories, consumables and spare parts Accessories Universal bracket 312.1160.0 Bracket for fastening the ATMOS S 201 Thorax on the patient bed, standard rail, wheelchair, tripod, etc. Consumables OT-Set 312.1031.0 In the surgical kit included: Canister for ATMOS S 201 Thorax...
  • Page 25 2l, conection to the device via direct docking, material: polycarbonate Suction hose ATMOS single-use suction container for Thorax, double lumen, with integrated bacterial filter in the measuring channel, 180 cm length Operating time...
  • Page 26: Checking / Reprocessing / Disposal

    Checking / Reprocessing / Disposal 10.1 Checking ATMOS suction devices For the ATMOS S 201 Thorax the calibration should be checked once a year and a safety-relevant check should be performed acc. to § 6 MPBetreibV. Depending on the number of recharging cycles of the batteries we recommend in this connection their exchange.
  • Page 27: Notes On Emc

    11.1 Guidelines and Manufacturer´s Declaration - Emissions The ATMOS S 201 Thorax is intended for use in the electromagnetic environment specified below. The customer or user of the ATMOS S 201 Thorax should ensure that it is used in such an environment.
  • Page 28 11.3 Guidelines and Manufacturer´s Declaration - Immunity The ATMOS S 201 Thorax is intended for use in the electromagnetic environment specified below. The customer or user of the ATMOS S 201 Thorax should ensure that it is used in such an environment.
  • Page 29 To determine the electromagnetic environment in regard to stationary transmitters, a study of the location is to be considered. If the measured field strength at the location where the ATMOS S 201 Thorax is used exceeds the above compliance level, the ATMOS S 201 Thorax is to be observed to verify the intended use.
  • Page 30 For your notes...
  • Page 31 For your notes...
  • Page 32 ATMOS MedizinTechnik GmbH & Co. KG in accepting such goods or payment arrears. Acts of God or stoppages upon in writing by us.

Table of Contents