5.0 Cleaning and care
5.1 General information on cleaning and disinfection
Prior to cleaning
Set main switch of the ATMOS
Prior to each use:
) The described measures relating to cleaning and
disinfection resp. sterilisation do not replace the relevant
instructions which must be adhered to prior to operation!
• For disinfection, you may use all surface and instrument
disinfectants listed in the following sub-chapter
"Recommended disinfectants".
• Please ensure that all surfaces are dry wiped. Use a
disposable cloth to absorb any liquid.
• If any disinfectant is spilled, make sure to dry wipe the
surface immediately in order to prevent the liquid from
penetrating gaps and edges.
5.1.1 Cleaning the unit surface
• The surfaces of the ATMOS
workstation are resistant against all surface disinfectants
listed in chapter „Recommended disinfectants".
• Wipe the unit surface with a cloth moistened with a
cleaning or disinfecting solution.
• Disinfectant sprays or disinfectant tissues may also be
used for cleaning and disinfection.
ª Please observe that the alcohol contained in these
agents could corrode or cloud the protective covers if
employed on a long term basis.
26
S 61 Servant ENT workstation to OFF prior to cleaning and disinfection!
®
S 61 Servant ENT
®
if
required
) Always observe the concentration specifications and
instructions by the respective manufacturer!
• Do not use
- Disinfectants which contain organic or inorganic acids
or bases as they could cause corrosion damage.
- Disinfectants containing chloramides or phenol
derivatives, since these may cause stress cracks in
the plastic material used.
• All application parts which are exposed to direct contact
with the patient during treatment must be exchanged
or cleaned and disinfected immediately for hygienic
reasons.
• Only deposit clean instruments on the board!
• Clean and disinfect the instruments regularly!