2.0 For your safety

2.1 For your safety
• The ATMOS
S 61 Servant ENT workstation is produced
®
according to IEC 601 / EN 60601 and listed in the
following classes:
- VDE Class of protection 1
- Class IIa (EEC 93/42).
• The device may only be connected to a properly installed
grounded electrical outlet.
• Correct configuration in assembly of country-specific
connections:
- green / yellow: protective conductor (PE)
- blue: neutral conductor (N)
- black or brown: phase (L)
• Caution! Mirror and endoscope heaters may generate
temperatures above 40°C!
• The ATMOS
S 61 Servant ENT workstation may only
®
be used under the supervision of skilled staff who have
been authorised by ATMOS and trained in its operation
(IEC 601-1 / EN 60601-1).
• The mains voltage indicated on the type plate must
correspond to the values of the supply network.
• Make sure prior to every application of the equipment that
it is technically safe and in proper condition. Damaged
leads and hoses must be replaced immediately!
• Display instruments and valves must be checked for
correct function at regular intervals!
• Every morning a visual inspection of the hoses, secretion
canister and cables must be performed. Exchange
damaged parts! Only proper and undamaged plugs and
extension cables may be used.
• Switch off the unit before you open it and separate it from
the line voltage.
• Maximum load for clipboard: 2 kg!
• Never leave the patient unattended at the treatment unit.
• The control panel must be clearly visible and accessible
for the user.
• This product is not re-sterilizable. Repeated reuse of
components which are marked with a
In case of repeated reuse these components lose their
function and there is a high infection risk.
• In case that there is no adapter attached to the light
conductor, touching the light source with either the fingers
or any tools should be avoided. On the one hand this
may damage the lenses and on the other hand there is a
risk of injury.
6
!
2
is forbidden.
• Existing water consuming devices may only be switched
on when the water supply is guaranteed!
• The ENT unit requires clean water (drinking-water
quality) for the operation. In case the clean water cannot
be provided by the water supply, a pre-filter must be
installed. The relevant country specific regulations for the
installation must be considered!
• Please note:
A medical insulating transformer with earth leakage
monitor or any similar safety system acc. to EN 60 601-1
is required, if several devices are connected over one
common power supply. The transformer must correspond
to the power consumption of all the devices to be
connected.
• Do not place used contaminated instruments on the ENT
unit except on destined places!
• The ambient conditions specified in the "Technical data"
must be strictly observed!
• Switch off main switch after finishing work in practice and
close water supply, if present.
• Care is to be paid in respect to light sources when
working with endoscopes. Do not look directly into the
light outlets! In case of possible light failure remove the
endoscope from the working area.
S 61 Servant ENT workstation may be
• The ATMOS
®
operated only in rooms used for medical purposes, but
not in areas subject to explosion hazards and in oxygen
rich environments.
• The ATMOS
S 61 Servant ENT workstation meets the
®
immunity to interference requirements of IEC 601-1-2 /
EN 60601-1-2 „Electromagnetic Compatibility – Medical
Electrical Devices".
• The ATMOS
S 61 Servant ENT workstation may
®
not be operated with devices not complying with the
requirements of standard EN 60601-1 „Medical Electrical
Equipment" and EN 60601-1-2 „Electromagnetic
Compatibility" (Medical Electrical Equipment).
• ATMOS is not liable for personal injury and damage to
property if
- no original ATMOS parts are being used,
- the advice for use in these operating instructions is
not being observed,
- assembly, new settings, alterations, extensions
and repairs have been carried out by personnel not
authorised by ATMOS.
• Do not restrict the air supply at the rear of the unit!
• Prior to starting the treatment application parts have to be
checked regarding their temperature.
• It is not allowed to use flammable substances with the
device.
• The system frame for 2 modules incl. microscope must
be screwed into the floor.