5.0 Cleaning and care

5.3 Cleaning method for handle with jet connection for water irrigation tip

The ATMOS devices Variotherm plus and Hygrotherm plus are properly used with disposable hose tips. These hose tips must
be disposed of after each application to the patient.
When using the hose tips, ATMOS recommends reprocessing according to the cleaning and disinfection plan below.
What
Parts to be reprocessed
Ear irrigation / Thermal nystagmus stimulation
Handle
Jet connection
Splash guard
Hose tip (dis-
posable)
Rinsing attach-
ment
Hygiene filter
Recommended disinfectants
Surface disinfection for coat-
3)
ed surfaces:
Green & Clean SK (ATMOS)
Dismozon
plus (Bode Chemie)
®
Kohrsolin
FF (Bode Chemie)
®
Perform
(Schülke & Mayr)
®
Terralin
Protect (Schülke & Mayr)
®
For concentrations, contact time, temperature and material compatibility, please see the relevant information from the manufacturer.
Important information
Wipe cleaning and disinfection: All surfaces must be wiped with a clean (disposable) wipe which is dampened with disinfectant solution; the
entire surface must be wiped thoroughly and may not be dried afterwards.
Please observe the manufacturer's instructions for use.
1)
Preferred: machine cleaning and disinfection in a washer disinfector
2)
Material dimensionally stable at 134 °C
6)
16
How
C
D
S
Cleaning
Disinfection Sterilization
X
X
3
X
X
2,4,5.6
X
X
2,4,5
X
X
2,4,5
Other surfaces:
Dismozon
plus (Bode Chemie)
®
Kohrsolin
FF (Bode Chemie)
®
Bacillocid
rasant (Bode Chemie)
®
Mikrobac
forte (Bode Chemie)
®
Perform
(Schülke & Mayr)
®
Terralin
Protect (Schülke & Mayr)
®
Surface disinfectant FD 312 (Dürr
Dental)
Quick disinfection B 30 (Orochemie)
Details
Wipe cleaning and disinfection
Cleaning and disinfection (manual or
automatic)
Cleaning and disinfection (manual or
automatic)
Change after each application
Cleaning and disinfection (manual or
automatic)
See operating instructions for hygiene
filter
Manual disinfection of instru-
4)
ments:
Korsolex
AF (Bode Chemie)
®
Korsolex
basic (Bode Chemie)
®
Korsolex
plus (Bode Chemie)
®
Korsolex
extra (Bode Chemie)
®
neodisher
Septo MED (Dr. Weigert)
®
neodisher
Septo 3000 (Dr. Weigert)
®
Sekusept
PLUS (Ecolab)
®
Sekusept
aktiv (Ecolab)
®
Gigasept
Instru AF (Schülke & Mayr)
®
Gigazyme
(Schülke & Mayr)
®
Gigasept FF neu (Schülke & Mayr)
When
Qualified and trained
staff who are familiar
with reprocessing
responsible person
-> use a water-based
overhead marker)
X
X
X
X
X
X
Automatic disinfection of instruments:
5)
Dismoclean
21 clean (Bode Chemie)
®
Dismoclean
24 Vario (Bode Chemie)
®
Dismoclean
28 alka one (Bode Chemie)
®
Dismoclean
twin basic/twin zyme (Bode Chemie)
®
neodisher
FA (Dr. Weigert)
®
neodisher
FA forte (Dr. Weigert)
®
neodisher
MediClean forte (Dr. Weigert)
®
Thermosept
alka clean forte (Schülke & Mayr)
®
Thermosept
RKN-zym (Schülke & Mayr)
®