3.0 Setting up and starting up

3.9 Adjustment of front feet

3.10 Connections in the service compartment

3.10.1 Hoses

Insert phone jack into marked jack
Insert the three-hole gasket at the
inlet hose of the water connection for
ear irrigation.
) A „three-hole gasket" must be used
otherwise a malfunction of the
device could be the result!
First start-up
Prior to shipment each ATMOS
safe functioning of the device after transport and installation, the following points should be observed:
The user should only operate the device when
1. a functional test was carried out on the device at the place of operation.
2. the operating instructions were read and acknowledged.
After transportation/storage of the device at low temperatures and prior to first start-up it should be kept at room temperature for
at least four hours. If the device is not acclimatised the formation of condensation water is possible and a malfunction might be
the result.
Adjust the screws of the feet through the boreholes with a flat-bladed screwdriver scaled
4-5.
Insert application parts into the hanging
strip
Connect water hose in the service
compartment
Open the tap on the treatment unit.
Check whether all connections are
leakproof.
S 61 Servant was checked by the manufacturer for function and safety. In order to ensure the
®
Connect compressed-air hose
in the service compartment
13