TASKS/TESTS
Remove the shipping hardware
Install the power supply
If applicable: Change Wrist Angle
If applicable: Change setting for 50-plate stacks
Install the BioStack with the interfacing instrument*
If applicable: Connect the host computer to the instruments, install
software, and test communication*
Run the System Test
Run the Checksum Test
Run the Instrument Alignment Utility**
Run the Verify Test**
If applicable: Install the BioStack barcode scanner and run the
Scanner Test‡
* Refer to the instrument-specific instructions.
** Perform the Instrument Alignment and Verify Test whenever the BioStack is
moved from one location to another.
Performance Qualification (PQ) procedures are not required for the BioStack.
n
Performing chemistry testing on the microplates is not applicable to the
BioStack.
‡
Install the Barcode Scanner on page 94
Recommended Qualification Schedule | 71
.
IQ
OQ
P
P
P
P
P
P
P
P
P
P
P
P If Verify
Test fails
P
P
P
BioTek Instruments, Inc.