68 | Chapter 4: Maintenance
 5. Apply a small amount of new lubricant to all surfaces of the linear ways,
spreading thin layers and working the grease into the surfaces with a
toothbrush.
 
Ensure that you lubricate the sides of the linear ways, since ball bearings
l
travel on these surfaces.
 6. Return the BioStack to its upright position and reconnect its power supply.
 7. Turn on the BioStack and run a Verify Test to confirm that moving the BioStack
did not compromise its alignment with the interfacing instrument. (Refer to the
instrument-specific PDF file.)
 
If the Verify Test passes, the BioStack is ready to resume normal operation.
l
 
If the Verify Test fails, realign the gripper with the plate carrier using the
l
Alignment Utility and rerun the Verify Test. If the test continues to fail,
contact BioTek's Technical Assistance Center.
BioStack™ Microplate Stacker