Probable Causes
2300
Gripper width saved is outside allowed range. Incorrect measurement
entered. Incorrect units.
2401
Hardware not present. A dual carrier is not present. An internal code error is
likely. Contact BioTek TAC.
2402
Hardware not present. An odd spaced dual carrier is not present. An internal
code error is likely. Contact BioTek TAC.
2403
Hardware not present. A wrist is not installed. An internal code error is likely.
Contact BioTek TAC.
2404
A barcode reader is not present.
make sure the setting is correct for the barcode scanner. If problems persist,
contact BioTek TAC.
2405
Alternative wrist position is not selected.
Requirements on page 26
angle matches your instrument requirement. Contact BioTek TAC if problems
persist.
2600
A gripper interface dimension is out-of-range at power up. Rerun the
Instrument Alignment Utility to clear the error.
2601
The instrument's gripper release position is invalid. Contact BioTek TAC.
2602
The instrument's gripper clear position is invalid. Contact BioTek TAC.
2603
The instrument's gripper grip position is invalid. Contact BioTek TAC.
2700
Error attempting to run the barcode scanner SET command. A response
returned from the scanner is invalid.
2701
A response from the barcode scanner while attempting the SET command
does not contain a valid barcode location. Valid numbers are 0–3.
2702
The barcode type returned by the scanner is not one of the four types
supported by the scanner.
on page 99
2703
Barcode type is not supported. One of the four default barcode types is not
supported by the scanner.
on page 99
2704
Error disabling start/stop character transmission. While attempting to tell the
scanner to disable the transmission of start and stop characters along with
the barcode value for the Codeabar barcode type, an error occurred.
2800
Out of range error. An attempt was made to set a configuration parameter,
using an out-of-range value.
See Dip Switch Settings on page 28
and make sure the dip switch setting for wrist
See Preparing and Attaching Barcode Labels
for required barcode types.
See Preparing and Attaching Barcode Labels
for required barcode types.
Communication Error Codes | 81
See Wrist Angle Setting
BioTek Instruments, Inc.
and

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the BioStack 4 and is the answer not in the manual?

Questions and answers

This manual is also suitable for:

Biostack seriesBiostack 3Biostack 2wr

Table of Contents

Save PDF