34 | Chapter 2: Installation
5. When RE-STACK? appears, select
6. Press the
Main Menu
405 TS and MultiFlo FX (Touch Screen Instruments) Only
Enable the BioStack to operate with the washer/dispenser:
1. Turn on the washer/dispenser and allow its System Test to complete.
2. At the Home screen, select
3. Select the
BioStack
4. Press
Get Basecode Version
you get an error message, check the cabling and restart both instruments.
Test Communication
(Optional) Use the Software Data Sheet on page 30 to record software version
n
and checksum information.
If software version/checksum information is displayed without error, the
n
Communication Tests completed successfully.
: From the main menu, select
Keypad
.
BIOSTACK
: At the Home screen, select
Touch screen
Communication is successful if the BioStack basecode software version (the first
number), interface definition version (the second number), and checksum are
displayed.
If communication is not successful, look up the error code:
Codes on page 74
BioStack™ Microplate Stacker
key.
Instrument > Next > BioStack
checkbox to show it is installed.
to ensure the instruments are communicating. If
. Contact BioTek if you cannot resolve the problem.
.
YES
UTILS > TESTS > CHKSUM/SOFTWARE >
Instrument > Next > BioStack
.
Communication Error
Need help?
Do you have a question about the BioStack 4 and is the answer not in the manual?
Questions and answers