Handling
7.7 Open the Axon-Bus Gripping Component in an Emergency
7.8 Bluetooth Function
Pressing the button in the charging receptacle and holding for more than four seconds with the Axon-Bus prosthet
ic system turned off activates the Bluetooth function of the connected AxonMaster 13E500. The blue LED lights up.
Once a connection to the PC has been successfully established, the LED flashes blue.
7.9 Beep Function
Action
Turn the Axon-Bus pros
thesis solution on
Push the button for one to
four seconds.
Turn the Axon-Bus pros
thesis solution off
Push the button for more
than one second.
Charge
the
Axon-Bus
prosthesis solution
Connect
the
charging
plug, charging begins.
Stop charging
Disconnect the charging
plug. Charging ends and
the Axon-Bus prosthesis
solution is switched off.
Empty battery indicator
Turn Bluetooth function
on (if the Axon-Bus pros
thesis solution was previ
ously switched off)
Push the button for more
than four seconds.
Emergency opening of
the Axon-Bus terminal
device
Push and hold the button
for about three seconds
until the Axon-Bus terminal
device opens.
7.10 Data transfer between the product and the PC
Product settings using the adjustment software can only be made via Bluetooth data transfer. For this purpose, a
Bluetooth wireless connection must be established between the product and the PC using the "60X5=* BionicLink
PC" Bluetooth adapter. The installation and use of the "60X5=* BionicLink PC" adapter are described in the
instructions for use included with the adapter.
16
This safety function allows the Axon-Bus gripping component to be opened regard
less of the control signals present.
1) With the Axon-Bus prosthetic system switched on, press the charging receptacle
button and hold for approximately three seconds until the Axon-Bus gripping
component begins to open. A pulsating beep sounds as it opens.
2) Releasing the button immediately stops the process of opening the Axon-Bus
gripping component and turns the entire Axon-Bus prosthetic system off.
Beep
Beep
1x long
2x short
Beep
Beep
3x short
1x short, 1x
long
Beep
Pulsating
AxonArm Ergo 12K501=*