System Voltage Failure; Hardware Configuration Error - GE Versana Balance Basic Service Manual

Table of Contents

Advertisement

Diagnostics/Troubleshooting

7-4-3 System voltage failure

7-4-4 Hardware configuration error

7-10
System view
Figure 7-10. System voltage failure
Cause: Hardware and/or voltage error detected.
What to do
1. Press OK and reboot the system.
2. if the problem persists, shut down the system, turn OFF the
circuit breaker and then reboot the system.
Figure 7-11. Hardware configuration error
Cause: Hardware(s) is not detected as valid system.
System is able to boot up but not able to scan.
What to do
1. Press OK and reboot the system.
2. If the problem persists, shut down the system, turn OFF the
circuit breaker and then reboot the system.
3. if the problem persists, shut down the system and analyze
the log files in Technical Support Mode.
Versana Balance – Basic Service Manual
5808768-100 English Rev.9