Program 5:
One-electrode control (1 electrode or 1 linear transducer)
The upper threshold is set using MyoSelect (scale D).
Openingandclosing
according to the control mode
of the System Electric Greifer or
System Electric Hand suitable for
one-electrode control (e.g. Sen-
sorHand Speed control mode 3).
MyoSelect
Switch-
scale C
0
1
2
3
4
10 sec.
5
12 sec.
5,9
14 sec.
Automatic switch-back deactivated; switch-back exclusively by impulse
6
switching.
Upon delivery, the automatic switch-back is deactivated. The scale C in the MyoSelect's display
is therefore set to 6.
13E205
Switching between gripping function and rotation is performed
by impulse switching, which means that the control signal must
quickly exceed the upper threshold and then quickly subside. See
section 5.2 "Switching". In addition, a time control feature can
be used to switch back to the gripping function. The switch-back
time is set using the 757T13 MyoSelect. The respective setting
can be read on scale C on the MyoSelect's display.
Rotation can be controlled both digitally and proportionally (see
Section 5.1.2).
Supination
Electrode: By quickly and
strongly beginning muscle
contraction
Linear transducer:
By quickly beginning to pull
on the linear transducer
Note:
To avoid damaging the transducer, the linear transducer
must not be pulled beyond the stop.
back
time
2 sec.
4 sec.
Insert
6 sec.
battery
8 sec.
Functions
OPEN /
CLOSE
Pronation
Electrode:
By slowly beginning muscle
contraction
Linear transducer:
By slowly beginning to pull on
the linear transducer
Impulse switching or
automatically after
2 –14 seconds
PRO / SUP
Impulse switching
Ottobock | 23