Program 3:
Safety Cocontraction
Control with 2 electrodes
MyoSelect
Switch-
scale C
0
1
2
3
4
10 sec.
5
12 sec.
5,9
14 sec.
6
Automatic switch-back deactivated; switch-back exclusively by safety
Cocontraction.
Upon delivery, the automatic switch-back is deactivated. The scale C in the MyoSelect's display
is therefore set to 6.
Program 4:
Control with 2 electrodes and a MyoBock switch
Switching from gripping function to rotation is performed by actuating and holding the switch.
Switch-back to the gripping function takes place as soon as you let go of the switch.
22 | Ottobock
Switching between gripping function and rotation is performed
with safety Cocontraction. See section 5.2 "Switching". In ad-
dition, a time control feature can be used to switch back to the
gripping function. The switch-back time is set using the 757T13
MyoSelect. The respective setting can be read on scale C on
the MyoSelect's display.
Rotation can be controlled both digitally and proportionally (see
Section 5.1.2).
back
time
2 sec.
4 sec.
Insert
6 sec.
battery
8 sec.
Insert
battery
Safety Cocontrac-
tion or automatically
after 2 –14 seconds
OPEN /
CLOSE
Safety Cocontraction
Let go of
switch
OPEN /
CLOSE
Actuate and
hold switch
PRO / SUP
PRO / SUP
13E205