Otto Bock 13E205 Instructions For Use Manual page 21

Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 15
Insert
battery
Program 2:
Cocontraction
Control with 2 electrodes
MyoSelect
Switch-
scale C
0
1
2
3
4
10 sec.
5
12 sec.
5,9
14 sec.
6
Automatic switch-back deactivated; switch-back exclusively by cocontraction.
Upon delivery, the automatic switch-back is deactivated. The scale C in the MyoSelect's display
is therefore set to 6.
13E205
Leftarm
prosthesis
Rightarm
prosthesis
Switching between gripping function and rotation is performed
with Cocontraction. See section 5.2 "Switching". In addition, a
time control feature can be used to switch back to the gripping
function. The switch-back time is set using the 757T13 Myo-
Select. The respective setting can be read on scale C on the
MyoSelect's display.
Rotation can be controlled both digitally and proportionally (see
Section 5.1.2).
back
time
2 sec.
4 sec.
Insert
6 sec.
battery
8 sec.
Upperthreshold
Rotation: Pronation
Hand: CLOSE
MyoSoft/ PAULA:
blue line
Rotation: Supination
Hand: OPEN
MyoSoft/ PAULA:
red line
Slowly beginning
muscle contraction
Thresholds A, B adjustable with MyoSelect
Quickly beginning
muscle contraction
A
Upperthreshold
Rotation: Supination
MyoSoft/ PAULA:
Rotation: Pronation
Hand: CLOSE
MyoSoft/ PAULA:
Cocontraction or
automatically after
2-14 seconds
OPEN/
CLOSE
Cocontraction
B
Hand: OPEN
red line
blue line
OPEN/
CLOSE
PRO/SUP
PRO/SUP
Ottobock | 21

Advertisement

Table of Contents
loading

Table of Contents