Checking The Vacuum Pressure; Figure 16. Ion Trap Page - Thermo Scientific ETD Getting Started Manual

For the ltq xl and ltq velos ms detectors (compatible with ltq orbitrap xl and ltq orbitrap velos)
Table of Contents

Advertisement

4
Daily Operation
Before Operating the LTQ XL/ETD or LTQ Velos/ETD System

Checking the Vacuum Pressure

26
MS/ETD System Getting Started Guide
Note
When on, the System LED on the front panel of the MS detector turns green and the
high voltage to the electron multipliers turns on.
The following conditions can cause the MS/ETD System to remain in standby mode
even though you try to turn it on:
• Attempting to turn on the reagent ion source when the restrictor heater, source
heater, and transfer line heater are not at their target temperatures.
• When either the LTQ XL/ETD or LTQ Velos/ETD MS detector or the
MS/ETD System goes into standby mode. Reagent vial nitrogen cooling turns on
if the vials are at an elevated temperature.
• Exception: If you put the LTQ XL/ETD or LTQ Velos/ETD system in standby
mode, the cooling nitrogen turns on after a one hour delay.
• Whenever the pressure in the LTQ XL or LTQ Velos MS detector or the
MS/ETD System exceeds its protection limit. Reagent vial nitrogen cooling turns
on if the vials are at an elevated temperature.
• Whenever the flow of reagent ions becomes insufficient as determined by the
AGC setting. When this occurs the LTQ XL/ETD or LTQ Velos/ETD system
completes the Xcalibur sequence step in progress before going into standby mode.
This prevents the loss of analysis results that might be affected by the reduced
reagent ion flow.
Check the vacuum pressure before operating the system.
To check the vacuum pressure
1. In the Tune Plus window, choose Setup > Vacuum

Figure 16. Ion Trap page

(Figure
16).
Thermo Scientific

Advertisement

Table of Contents
loading

Table of Contents