Figure 14. Semi-Automatic Page In The Tune Dialog Box - Thermo Scientific ETD Getting Started Manual

For the ltq xl and ltq velos ms detectors (compatible with ltq orbitrap xl and ltq orbitrap velos)
Table of Contents

Advertisement

Thermo Scientific
6. Click the Semi-Automatic tab

Figure 14. Semi-Automatic page in the Tune dialog box

7. Under What to Optimize, select one of the following items to tune:
• Reagent ion lens 1 (V)
• Reagent ion gate lens (V)
• Reagent ion lens 2 (V)
• Reagent ion lens 3 (V)
• Back Multipole Offset (V)
• Back lens (V)
8. Under Optimization Range, adjust the settings in the Start, End, and Step boxes.
9. Click Start.
(Figure
14).
MS/ETD System Getting Started Guide
3
Tuning
Tuning the Reagent Ion Optics
23