Figure 13. Graph View Of The Reagent Ion Source Tuning In The Tune Plus Window - Thermo Scientific ETD Getting Started Manual

For the ltq xl and ltq velos ms detectors (compatible with ltq orbitrap xl and ltq orbitrap velos)
Table of Contents

Advertisement

Figure 13. Graph view of the reagent ion source tuning in the Tune Plus window for the LTQ XL
Thermo Scientific
8. Select the Reagent Ion from Vial 1 or Reagent Ion from Vial 2 check box, whichever
appears and is available.
9. Click Start.
A plot of the reagent ion intensity appears
• The response of the reagent ion intensity to changes in the lens parameters in the
Reagent Ion Optics dialog box
• The emission current, CI gas pressure, and electron energy in the Reagent Ion Source
dialog box
(Figure 8
Adjust these parameters to achieve the maximum reagent ion signal intensity.
(Figure
(Figure 11
on
on
page
16)
Tuning the Reagent Ion Optics
13). You can observe:
page
20)
MS/ETD System Getting Started Guide
3
Tuning
21

Advertisement

Table of Contents
loading