Software Configuration; Testing The Connection; Checking Wireless Signal Strength - Thermo Scientific Smart-Vue Installation Manual

C02/temperature module
Hide thumbs Also See for Smart-Vue:
Table of Contents

Advertisement

Software
configuration

Testing the connection

Checking wireless signal
16
Smart-Vue CO
and Temperature Module Installation Guide
2
When you activate your Smart-Vue end-point module as described in the
previous section, the device adds itself to your current Smart-Vue
Server/Smart-Vue Client configuration.
You may use Smart-Vue Client to test the wireless signal strength as
described here, entering your login name and password when prompted by
strength
the software.
1. Launch Smart-Vue Client on your computer and enter your login
name and password to continue.
2. Select Settings ==> Sensor settings (
3. Click on Add/Update an end-point module (or press F11 again).
4. Select your receiver in the drop-down menu ("local" as shown below
in Figure 11), and then select the wireless end-point module from
the sensor list on the left.
Figure 11. Configuration with Smart-Vue CO
5. Click on Power to test the connection. Click on Initialize, Current
sensor measurement, and Synchronize end-point module clock to
confirm proper operation. Click on Close to exit.
If this test does not show a wireless signal, please refer to the Thermo
Scientific Smart-Vue Software User Manual for instructions on adding
end-point modules into the system manually.
module on receiver "local"

Software configuration

), or press F11.
/temperature end-point
2
Thermo Fisher Scientific

Advertisement

Table of Contents
loading

Table of Contents