Daily Operation
Turning On the Reagent Ion Source
Figure 19. Fluoranthene radical anion mass spectrum
30
MS/ETD System Getting Started Guide
7. Check the spectrum and do one of the following:
• If the spectrum is satisfactory, proceed to
• If the spectrum is not satisfactory, tune the reagent ion optics as described in
Chapter 3, "Tuning."
8. In the Reagent Ion Source dialog box, clear the View Reagent Ion Spectra check box.
This action clears the Tune Plus spectrum view from the ETD module.
step
8.