Thermo Scientific ETD Getting Started Manual page 66

For the ltq xl and ltq velos ms detectors (compatible with ltq orbitrap xl and ltq orbitrap velos)
Table of Contents

Advertisement

6
Running an Xcalibur Experiment by Using ETD
3
Data Dependent NL MS
Experiment
50
MS/ETD System Getting Started Guide
12. In the Isolation Width (m/z) box, enter a value between 2 and 3.
The Normalized Collision Energy and Activation Q parameters are unavailable because
they are not applicable to ETD.
13. In the Activation Time (ms) box, leave the default value or enter a different value as
described in
"Optimizing the Reagent Ion Reaction Time"
14. In the Tune Plus window, click the Save
15. Browse to the
:\Xcalibur\methods folder, and specify a file name.
16. Click Save.
on
page
(
) button.
38.
Thermo Scientific