Troubleshooting
Thermo Scientific
The MS/ETD System uses the following consumables:
• Filament
• Fluoranthene
• Ion volume
After a period of time, you might need to replenish the ETD fluoranthene reagent, replace the
filament, and clean or replace the ion volume. As the consumables are depleted, the ETD
reagent m/z peak intensity diminishes (m/z = 202 in negative polarity mode). Periodically
checking this peak intensity is a good way to monitor the ETD module consumables. To view
the ETD reagent m/z signal, follow the procedure in
page
28.
Tip Expect the filament and ion volume to last for about 400 to 500 hours of operation,
and a vial of fluoranthene reagent to last for several months or more.
Table 3
lists some ETD module problems, their causes, and their possible solutions.
"Turning On the Reagent Ion Source"
MS/ETD System Getting Started Guide
7
on
51