Data Securities
The N9322C offers the optional memory erase function for
data security. Press [System] > {More} > {Securities} > {Erase
Memory} to erase all the user data in internal memory. Press
Enter as a terminator to start the erase process immediately.
The memory erase process takes about 15 minutes. During
the time, there must be a constant power supply to ensure
the successful erase. If the erase process is interrupted,
please reboot the instrument and erase memory again.
Probe Power Output
The Probe Power provides power for high- impedance AC
probes or other accessories (+15 V, –12V, 150 mA maximum).
The Porbe Power is set to off as default. Press [System] >
{More} > {Port Setting} > {Probe Power On} to switch on the
porbe power output.
Getting Started
Helpful Tips
27