Spectrum Emission Mask - Keysight Technologies N9322C User Manual

Spectrum analyzers
Hide thumbs Also See for N9322C:
Table of Contents

Advertisement

Key Reference
Meas

Spectrum Emission Mask

Avg Number
Ref Channel
124
Spectrum Emission Mask measurement includes the in- band
and out- of- band spurious emissions. As it applies to
W- CDMA (3GPP), this is the power contained in a specified
frequency bandwidth at certain offsets relative to the total
carrier power. It may also be expressed as a ratio of power
spectral densities between the carrier and the specified
offset frequency band. For WLAN, the reference power is
taken as the peak PSD in the signal, and all offset results
are also the peak PSD in that offset as opposed to the
integrated power.
Set Avg Number to On to specify the number of measurement
On Off
averages used when calculating the measurement result. The
average will be displayed at the end of each sweep. Setting
Avg Number to Off disables measurement averaging.
Key access: Meas > SEM
Meas Type
Displays a menu to select a measurement reference type,
Total Pwr Ref or PSD Ref.
Key access: Meas > SEM
Total Pwr Ref
PSD Ref
Accesses the softkeys to change the reference channel
settings.
Key access: Meas > SEM
Chan Integ BW Specifies the integration bandwidth used in
Chan Span
When Total Pwr Ref is selected, the power in the
carrier is used as the reference in computing
the relative power values for the offsets.
When PSD Ref is selected, the mean power
spectral density of the carrier is used in
computing the relative power spectral density
values for the offsets.
calculating the power in the main channel.
Specifies the span used in measuring the
power in the main channel.