Functions and Measurements
Making a Power Measurement with USB Power Sensor
Delay
Set the delay time to be applied between the trigger
event and all the gate start times. This allows you to
time- shift all the gates by the same amount with one setting
change.
Slope
Select the positive or negative slope type to
determine if the trigger event is recognized on the rising or
falling edge of a signal respectively.
Holdoff
Set the holdoff time to disable the trigger
mechanism after a trigger event occurs.
Hysteresis
Set the hysteresis to help generate a more stable
trigger by preventing triggering unless the RF power level
achieves the trigger level and the additional hysteresis value.
It can be applied to both rising and falling edge trigger
generation.
Hysteresis is only available for the internal trigger source
and manual trigger level.