Keysight Technologies N9322C User Manual page 184

Spectrum analyzers
Hide thumbs Also See for N9322C:
Table of Contents

Advertisement

Instrument Messages
Device-Specific Errors
176
767
Fatal LDS error
Attempt to initialize the instrument state has failed. Cycle
instrument power. If this fails to correct the problem,
contact your nearest Keysight Technologies service center.
768
Unable to load user state
Attempt to load a state failed, because the state was
saved with a
loaded. Load the appropriate personality and try again.
770
Incorrect filename
You sent a remote command to save a screen file but did not
specify a valid extension.
780
No Peak Found
No signal peak was found that meets the criteria under
Peak Search > Peak Criteria.
781
Video Trigger cannot be active
You sent a remote command to do one of the following:
• Turn on video trigger while the Average Detector or a
Marker Function is active.
• Turn on the Average Detector or a Marker Function while
Video Trigger is active.
782
Video Trigger cannot be used with Mkr Func
Due to the Average Detector, the video Trigger cannot be
used with Maker function.
783
Allowable span for current CF exceeded
784
Allowable CF for current span exceeded
805
Doing Alignment
measurement personality that is not currently