Functions and Measurements
Measuring Mul tiple Signals
40
If the signal peak is not present on the display, increase the
frequency span out to 20 MHz, turn signal tracking on,
decrease the span back to 2 MHz and turn signal tracking
off:
Press Span > 20 > MHz
Press Peak Search
Press Span > 2 > MHz
Figure 3-5 Unresolved Signals of Equal Amplitude
4
Change the resolution bandwidth (RBW) to 100 kHz so that
the RBW setting is less than or equal to the frequency
separation of the two signals:
Press BW > 100 > kHz
Notice that the peak of the signal has become flattened
indicating that two signals are present.
5
Decrease the video bandwidth to 3 kHz:
Press BW > Video BW > 3 > kHz
Two signals are now visible as shown in
front- panel knob or arrow keys to further reduce the
resolution bandwidth and better resolve the signals.
Figure 3-6
. Use the