Keysight Technologies N9322C User Manual page 14

Spectrum analyzers
Hide thumbs Also See for N9322C:
Table of Contents

Advertisement

Overview
Front Panel Overview
6
11
RF IN connector recieves the signal input. The maximum
damage level is average continuous power +33 dBm or DC
voltage 50 VDC. The impedance is 50 W. (N- type female).
12
PROBE POWER connector provides power for high- impedance
AC probes or other accessories (+15 V, –12 V, 150 mA
maximum).
13
CAL OUT connector provides an amplitude reference signal
output of 40 MHz at –25 dBm (BNC female).
14
TG SOURCE connector N- type female, is the source output for
the built- in tracking generator. The maximum damage level
is average continuous power +30 dBm or DC voltage 50 VDC.
The impedance is 50 W. (for Option TG7)
15
Standby Switch switches on all functions of the analyzer. To
switch the analyzer off, press the switch for at least 2
seconds. This deactivates all the functions but retains power
to internal circuits so long as the analyzer is connected to
line power.
16
On LED (green) lights when the analyzer is switched on.
17
Standby LED lights when the analyzer is connected to the
line power.
18
Remote LED lights when the analyzer is remotely controlled
by a PC via the USB host interface on the rear panel.
19
USB Device Connector provides a connection between external
USB devices and the analyzer, such as a USB memory
device.

Hide quick links: