652
Connect CAL OUT to RF IN
you must connect the CAL OUT to RF IN with the appropriate
cable.
751
Instrument state set to initial values
While trying to load a trace or state, the state information
was found to be in error. This may be because the state had
been stored on a later revision of analyzer firmware. A
default set of state variables was loaded instead. There is
nothing wrong with the analyzer.
752
Unable to load state from file
Attempt to load a state from a file failed.
753
Unable to save state to file
Attempt to save a state to a file failed. See the associated
error messages for the cause.
757
Unable to load user state
A user preset failed, so the factory preset values were used.
Save a valid state into user preset and try again.
758
Unable to save user state
Attempt to save a user preset state failed.
760
Unable to query state
Query of state over the remote interface was unsuccessful.
762
Unable to load file
A failure occurred while loading a file.
764
Unable to save file
Attempt to save a file failed; the file was not saved.
Instrument Messages
Device-Specific Errors
175