Sweep Time
Auto Man
Res BW
Auto Man
Total Pwr Ref
(PSD Ref)
Auto Man
Offset/Limits
Displays the menus to change the following parameters for
offset frequency settings and pass/fail tests: Offset, Start Freq,
Stop Freq, Sweep Time, Res BW, and Limits. Refer to the tables
below for default settings of particular radio standards of
interest.
Key access: Meas > Spectrum Emission Mask
Specifies the sweep time used in measuring
the power in the main channel.
Specifies the resolution bandwidth used in
measuring and the power in the main channel.
Displays the type of power measurement
reference selected by Measure Type.
Total Pwr Ref is the power in the carrier that is
used as the reference in computing the
relative power values for the offsets. When
Man is selected, this can be set by the user.
When Auto is selected, this is the measured
power in the reference carrier.
PSD Ref is the mean power spectral density in
the carrier that is used as the reference in
computing the relative power spectral density
values for the offsets. When Man is selected,
this can be set by the user. When Auto is
selected, this is the measured mean power
spectral density in the reference carrier.
Key access: Meas > Spectrum Emission Mask > Ref
Channel
Key Reference
Meas
125