Service Setup
The following window opens displaying the firmware version on the device
and SD card:
2.
Press Enter to update the firmware.
While the update is in progress, the following message is displayed:
3.
Press Enter to select the OK button.
The following message is displayed:
• If the update was not successful, the system gives corresponding error
• If the update was successful, the system shuts down.
Formating the Flash Memory
The Format Flash procedure is used to resolve memory or data corruption problems. It
returns the device to its factory settings by erasing all the data, including ECG records,
system configurations, device configurations, option codes, log files, audit files, patient
lists, and exercise protocols. Before attempting to format the device's flash memory,
68
messages.
a.
Press OK on the error message to return to the update screen.
b.
Check the files in the SD card.
MAC™ 2000 ECG Analysis System
2053535-003C