GE MAC 2000 Service Manual page 58

Ecg analysis system
Hide thumbs Also See for MAC 2000:
Table of Contents

Advertisement

Service Setup
Then the results of the test are displayed.
• The test passes if the following message is displayed in the window: Passed.
• The test fails if the following message is displayed in the window: Failed.
5.
When the test is done, press Cancel to close the results window.
Testing the USB Port
Use the USB Test to test the USB port.
1.
Open the Diagnostic Tests window as described in
page
2.
Connect a USB keyboard to the USB port on the rear panel of the device.
NOTE:
3.
Select USB Test.
The following window opens:
4.
Press any key on the USB keyboard and verify pass or fail:
• If the character that appears in the Character Input field matches the key
• If the character does not match the key you pressed, or no character appears
58
If the Internal Modem Test fails, do the following:
a.
Replace the internal modem as described in
Modem" on page
120.
b.
Rerun the Internal Modem Test.
c.
If the test still fails, replace the PWA as described
, and enable the modem option.
46.
The USB keyboard used for this test must match the language that is
selected in setup.
you pressed, the test passed.
in the Character Input field, the test failed.
MAC™ 2000 ECG Analysis System
"Replacing the Internal
in"PWA" on page 120
"System Diagnostics" on
2053535-003C