Ecg analysis system
Hide thumbs Also See for MAC 2000:
Table of Contents

Advertisement

2.
Position the printer motor on the printer module assembly and secure with the
two fasteners, as shown in the following photograph.
3.
Position the printer motor cover over the printer motor and secure with its
fastener, as shown in the following photograph.
4.
Reattach the printer module assembly to the bottom cover.
See
5.
Continue with the appropriate functional checkout procedure for this FRU.
See
Paper Sensor
The following procedures describe how to remove and replace the paper sensors.
Removing the Paper Sensor
Use the following procedure to remove the paper sensor from the print module
assembly.
1.
Before beginning any steps, follow the instructions in
FRU Replacement" on page
2.
Remove the printer module assembly.
See
2053535-003C
"Replacing the Printer Module Assembly" on page 95
"Functional Checkout" on page 129
"Removing the Printer Module Assembly" on page 94
MAC™ 2000 ECG Analysis System
for more information.
73.
Maintenance
for instructions.
"Preparing the System for
for instructions.
97