GE MAC 2000 Service Manual page 23

Ecg analysis system
Hide thumbs Also See for MAC 2000:
Table of Contents

Advertisement

Symbol Descriptions (cont'd.)
Symbol
2053535-003C
Description
Consult Instructions for Use
Consult the operating instructions.
Defibrillation-proof Type BF Applied Part
Identifies a defibrillation-proof type BF applied part on medical
equipment that complies with IEC 60601–1.
This device meets the requirements for protection against electric
shock for an earth-free (floating) applied part (one intended for contact
with patients).
Defibrillation-proof Type CF Applied Part
Identifies a defibrillation-proof type CF applied part on medical
equipment that complies with IEC 60601–1.
This device meets the requirements for protection against electric
shock for an earth-free (floating) applied part (one intended for contact
with patients) for cardiac application.
No User– or Field-serviceable Parts
Do not open or disassemble the device for any reason.
Protective Earth (ground)
Identifies the terminal of a protective earth (ground) electrode or any
terminal that is intended for connection to an external conductor for
protection against electric shock in case of a fault.
Non-ionizing Electromagnetic Radiation
Indicates that the equipment emits elevated, potentially hazardous,
levels of non-ionizing radiation (electromagnetic energy) for diagnosis
or treatment.
Follow Instructions For Use
Read and understand the operator's manual before using the device
or product.
As a mandatory action sign, this symbol is identified by a blue
background and white symbol.
CAUTION:
SAFETY GROUND PRECAUTION
Pulling on the cable can cause the cord to deteriorate resulting in
electrical problems.
Remove the power cord from the mains source by grasping the
plug. DO NOT pull on the cable.
CAUTION:
CONSULT ACCOMPANYING DOCUMENTS
There may be specific warnings or precautions associated with the
device that are not otherwise found on the label.
Consult the accompanying documentation for more information
about safely using this device.
MAC™ 2000 ECG Analysis System