GE MAC 2000 Service Manual page 168

Ecg analysis system
Hide thumbs Also See for MAC 2000:
Table of Contents

Advertisement

Electromagnetic Compatibility (EMC)
EMC Immunity Test
Immunity Test
Conducted RF
168
Compliance
Compliance
Test Level
Level
3 Vrms
Multilink Lead
150 KHz to 80
set: 3.0 Vrms
MHz
Value Lead set:
@ 2 Hz mod.
3.0 Vrms
KISS Lead set:
0.3 Vrms
MAC™ 2000 ECG Analysis System
Electromagnetic
Environment–Guidance
Do not use portable or mobile
RF communications equipment
closer to any part of the system,
including the cables, than
the recommended separation
distance calculated for the
equation applicable to the
frequency of the transmitter.
Recommended separation
distance:
Multilink Lead Set:
d = 1.2√P
Value Lead Set:
d = 1.2√P
KISS Lead Set:
d = 11.6√P
2053535-003C