Testing the Acquisition Module
Use the Acquisition Module Test to determine if the acquisition board is working
properly.
1.
Open the Diagnostic Tests window as described in
page
2.
Select Acquisition Module Test.
A window similar to the one shown in the following screen opens.
NOTE:
3.
Note the test result and press Cancel.
4.
If the result of the Acquisition Module Test was Failed, replace the main PWA
as described in
Checking the Battery Status
Use the Battery Test to determine the status of the Lithium-Ion battery. You must
perform this test while running on battery power.
1.
Disconnect the device from AC power and wait at least one minute.
2.
Open the Diagnostic Tests window as described in
page
3.
Select Battery Test.
A window similar to the one shown in the following screen opens.
2053535-003C
46.
The version displayed depends on the software installed.
"PWA" on page
46.
MAC™ 2000 ECG Analysis System
"System Diagnostics" on
120.
"System Diagnostics" on
Service Setup
51