GE MAC 2000 Service Manual page 83

Ecg analysis system
Hide thumbs Also See for MAC 2000:
Table of Contents

Advertisement

3.
Remove the power supply assembly.
See step
91
4.
If it is installed, remove the KISS pump and tubing.
See step
for instructions.
5.
Remove the fasteners securing the power supply/KISS pump base to the chassis,
as identified in the following photograph.
6.
Lift the power supply/KISS pump base from the chassis.
2053535-003C
3
through step
6
for instructions.
2
through step
5
MAC™ 2000 ECG Analysis System
in
"Removing the Power Supply Assembly" on page
in
"Removing the KISS Pump Assembly" on page 126
Maintenance
83