Maintenance
Replacing the Printhead
Use the following procedure to replace the print head in the printer module assembly.
1.
Hold the new print head so the longer ribbon cable is on the right, as shown
in the following photograph.
2.
With the front of the printer module assembly facing you, gently thread the
printer ribbon cables through the printer module assembly and snap the print
head into the metallic print head holder, as shown in the following photograph.
NOTE:
3.
With the rear of the printer module assembly facing you, secure the print head
to the printer module assembly using the two M3x4 fasteners, as identified
in the following photograph.
4.
Reattach the printer module assembly to the bottom cover.
See
5.
Continue with the appropriate functional checkout procedure for this FRU.
See
102
Make sure the longer printer ribbon cable is on the side of the printer module
assembly with the lever and not the printer motor. When facing the front of
the printer module assembly, this would be on the right.
"Replacing the Printer Module Assembly" on page 95
"Functional Checkout" on page 129
MAC™ 2000 ECG Analysis System
for instructions.
for more information.
2053535-003C