Maintenance
3.
Connect the cables and harnesses, identified in the following photograph,
to the PWA.
4.
Position the power supply/KISS pump base on the chassis.
5.
Secure the power supply/KISS pump base to the chassis using its fasteners, as
identified in the following photograph.
6.
If it was installed, replace the KISS pump and tubing.
See step
for instructions.
7.
Replace the power supply assembly.
See step
92
8.
Reattach the top cover assembly.
See
instructions.
9.
Reinsert the battery.
See step
10.
Continue with the appropriate functional checkout procedure for this FRU.
See
88
1
through step
4
1
through step
4
for instructions.
"Reattaching the Top Cover Assembly to the Device" on page 81
6
through step
8
"Functional Checkout" on page 129
MAC™ 2000 ECG Analysis System
in
"Replacing the KISS Pump Assembly" on page 127
in
"Replacing the Power Supply Assembly" on page
in
"Replacing the Battery" on page 74
for more information.
for
for instructions.
2053535-003C