GE MAC 2000 Service Manual page 140

Ecg analysis system
Hide thumbs Also See for MAC 2000:
Table of Contents

Advertisement

Troubleshooting
Printer Internal Error Codes (cont'd.)
Error
Code
5
6
7
NOTE:
• Continuous printing of high amplitude noisy waveforms while in battery mode may
• Printing highly noisy waveforms at low speed for long periods of time may trigger
140
Cause
Printer driver timeout error
Printer driver miscellaneous
error
Undefined printer status was
received
cause the printing to stop and trigger the message Printing not possible Battery
Low.
If this happens, check the signal quality, connect the device to the mains or reduce
the gain, and restart printing.
the message: Printer internal error 2.
This indicates the printhead temperature exceeded operating limits and caused
the printing to stop.
If this happens, adjust the patient lead connections to remove any noise, check
the signal quality, and restart printing.
MAC™ 2000 ECG Analysis System
Solution
1.
Check the cables and replace if necessary.
2.
Replace the main PWA.
1.
Check the cables and replace if necessary.
2.
Replace the main PWA.
1.
Check the cables and replace if necessary.
2.
Replace the main PWA.
2053535-003C

Hide quick links: