Maintenance
4.
Turn over the top cover, remove the two M3x5 fasteners, and lift the deco, as
shown in the following photographs.
5.
Turn over the top cover and press down on the rear of the printer door release
lever, as shown in the following photograph.
6.
Grasp both sides of the raised printer door release lever, as shown in the
following photograph, and lift to remove it from the top cover.
7.
Grasp both sides of the plunger and lift to remove it from the top cover, as
shown in the following photograph.
You now have removed the top cover from the top cover assembly, as shown
in the following photograph.
78
MAC™ 2000 ECG Analysis System
2053535-003C